Description
Flamephotometer
Microprocessor Flame Photometer
FEATURES Extremely useful for medical and agriculture laboratory use All element measurement in single aspiration of the sample Automatic Optical Management of filters Calibration curve programmability using maximum of 5 standards.
Product Overview
Data processing through curve fitting techniques – least squares, quadratic Multiple calibration curves saving facility Samples data storage facility Printer Attachment Facility Measures in ppm, mg, % and meq Help key available at each and every operating step Specifications Range (ppm) Na: 0-100 K: 0-100 Ca: 0-100 (meq) Up to 250 meq/l, 1:100 dil Up to 250 meq/l, 1:100 dil Up to 250 meq/l, 1:100 dil Sensitivity Na:0.1ppm, K:0.01 ppm, Ca:0.5 ppm Units ppm, mg and % age Selectable Filter Selection Na & K and Ca by Microprocessor controlled stepper motor Reproducibility ±1% of FS ± 1 Digits Curve fit accuracy ± 2% of FS Display Large 20 x 4 Character LCD display with backklit Aspiration time 4-5 Seconds Flame system LPG & Dry oil free air Detector Photodiode Calibration Up to 5 point calibration with curve fitting software Gas Control Adjustable with knob Ignition system Auto ignition Fuel Gas LPG Burner Stainiless Steel Dimensions 365 X 245 X 220mm Approx. (LxBxH) Printer Provision…
Technical Specifications
| Range | (ppm) Na: 0-100 K: 0-100 Ca: 0-100 — (meq) Up to 250 meq/l, 1:100 dil Up to 250 meq/l, 1:100 dil Up to 250 meq/l, 1:100 dil |
|---|---|
| Sensitivity | Na:0.1ppm, K:0.01 ppm, Ca:0.5 ppm |
| Units | ppm, mg and % age Selectable |
| Filter Selection | Na & K and Ca by Microprocessor controlled stepper motor |
| Reproducibility | ±1% of FS ± 1 Digits |
| Curve fit accuracy | ± 2% of FS |
| Display | Large 20 x 4 Character LCD display with backklit |
| Aspiration time | 4-5 Seconds |
| Flame system | LPG & Dry oil free air |
| Detector | Photodiode |
| Calibration | Up to 5 point calibration with curve fitting software |
| Gas Control | Adjustable with knob |
| Ignition system | Auto ignition |
| Fuel Gas | LPG |
| Burner | Stainiless Steel |
| Dimensions | 365 X 245 X 220mm Approx. (LxBxH) |
| Printer | Provision for attachment of Dot-Matrix Printer or HP LaserJet / inkjet printer (non-win printers) with centronics parallel interface. |
| RS 232 interface | Available (PC software at extra cost) |
| Air Compressor | with Built-in Regulator and Air Filter to deliver stable and moisture/oil free air supply @ 1Kg/cm2 |
| Air Supply | By oil free Mini Compressor Unit With Pressure Regulator |
| Weight | Main Unit : 8 Kg. (Approx) compressor : 6.5 kg. (Approx) |
| Power Supply | 95 to 270V A 50-60Hz with inbuilt stabilizer |
| Accessories | Compressor with built-in air regulator and air filter to deliver stable and moisture free air supply @ 1Kg/cm22, 1mtr Capillary Tube, 2 mtr Gas Connector Tube, 0.5KVA Stabilizer already in-built in the system. |
Features and Details
- Extremely useful for medical and agriculture laboratory use
- All element measurement in single aspiration of the sample
- Automatic Optical Management of filters
- Calibration curve programmability using maximum of 5 standards.
- Data processing through curve fitting techniques – least squares, quadratic
- Multiple calibration curves saving facility
- Samples data storage facility
- Printer Attachment Facility
- Measures in ppm, mg, % and meq
- Help key available at each and every operating step
About Interlabs
Interlabs, associated with International Biological Laboratories, supplies scientific, laboratory, educational and engineering equipment from India. The company identifies its history as dating to 1958. Contact Interlabs to confirm product details and supporting documentation for your requirement.
Request Product Information
Share your application and technical requirements for Microprocessor Flame Photometer to discuss suitability, available configurations and a quotation.
Frequently Asked Questions
What is Microprocessor Flame Photometer used for?
FEATURES Extremely useful for medical and agriculture laboratory use All element measurement in single aspiration of the sample Automatic Optical Management of filters Calibration curve programmability using maximum of 5 standards.
What should I confirm before ordering Microprocessor Flame Photometer?
Confirm the intended application, technical specifications, available configuration, accessories, documentation and delivery requirements with Interlabs.